16 channel switch diags fail with 2247385 due to a design change in diode biasing made to recently released preamps with…

16 channel switch diags fail with 2247385 due to a design change in diode biasing made to recently released preamps with part number 5400044.

Symptom

16 channel switch diags fail with 2247385 Test: ''16 Ch Switch Input Osc. LB(L)'' diagnostic error The noise energy level is too high for channels 1-8 .

Error message

2247385 Test: ''16 Ch Switch Input Osc. LB(L)'' diagnostic error The noise energy level is too high for channels 1-8 .

2247387 The noise energy level is lower than expected for channels: 2 Thu Feb 24 00:34:59 2005 2247387 2 lx-mr5 diagCoord rrfRxFruIso.c 1539 ^MDiagnostic: 16 Channel Switch Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 10509 Object: APS Test Task Test: '16 Ch Switch Input Osc. LB, Port L' diagnostic error. The noise energy level is lower than expected for channels: 2->

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution