Unable to scan with HNS coil with error code 2218701 logged due to Tech had plugged HNS coil into wrong LPCA A/B port.

Symptom

Unable to scan with HNS coil with error code 2218701 logged See Long Description See below for actual error from site. Sun Jun 05 10:27:05 2011 2218701 scn ../RxMgrDownload.m 579 Exception Class: Unknown Function: Host: Scan Error while filling coilInfo of prescribed coil, see the log for details. Download failed .

Error message

2218701 Error while filling coilInfo of prescribed coil, see the log for details. Download failed. See Long Description 2218701 scn ../RxMgrDownload.m 579 Exception Class: Unknown Function: Host: Scan Error while filling coilInfo of prescribed coil, see the log for details. Download failed.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR