Coil ID light does not light up for any coil plugged into one port, but no errors in log, due to Port connector screw ha…

Coil ID light does not light up for any coil plugged into one port, but no errors in log, due to Port connector screw had backed partially out, blocking connection

Symptom

Coil ID light does not light up for any coil plugged into one coil port, with no errors in log See Long Description When plugging in coil in this port, seems more physical resistance than usual, then once plugged in the coil seems loose. The coil ID light would not come on, yet no errors in log.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution