LightSpeed 4.X - GRE
Symptom
Scan hardware stopped scan with DAS Board errors. Scan hardware stopped scan DAS Board errors Scanner Hardware Stopped Scan message appears in the operator console area during Axial Scan data acquisition with Error: 260001023 The DCB has notified the system of an error condition which prevents scanning and error: 260007402-Converter Board(s) Asserted a Board FAULT cards 76,78,80,82,84,86,88,90,92,94,96 and Error #260007774 list of converter cards from previous error message Problem was intermittent Then after time the errors would occur more often and eventually the problem was consistent [insite@erct ~]$ /usr/g/insite/bin/viewlog Sun Aug 29 Error: 260001023 Host: OBC Process: HOP File: hop erlog.c Line: 367 Function: No System Function Reported Scan: 50000/1/2 Type: Axial Exception Level: Pri/Most Severe Time: 08:37:10:145 Log Series: 120 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260002358 Notification level: 5 Sequence number: 11
Error message
Error: 260007774 List of Converter Cards From the Previous Error Message Card List: xx xx See Long Description for full error Tue Dec 30 09:25:17 2008 Error: 260007774 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 852 Tue Dec 30 09:25:17 2008 Host : Das Ermes # : 260007774 Exception Class : Secondary Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 852 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 List of Converter Cards From the Previous Error Message Card List: 8,11,14-15,19,21,25-27,29-36,38-58,60-63 Note on this error. Sometimes the actual bad board will be the one that is not listed.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2371627-4 replaces 2371627-3 GDAS / MDAS 16 Slice RIGHT Backplane Chassis Asm EDITION 1-2 See LD for FEMC comments 2371627-4 replaces 2371627-3 GDAS / MDAS 16 Slice RIGHT Backplane Chassis Asm EDITION 1-2 Part #: 2371627-4 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS-MDAS 16 Slice RIGHT BP-Chassis Asm EDITION 1-2 FRU: Yes Status: Orderable Comments: Added as replacement for 2371627, -2, -3 & 2332630 per ecr 2012678, ms13452 050518 Part #: 2371627-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: BP-Chassis Right EMC Asm - GDAS FRU: Yes Status: Orderable Comments: 2371627-3 is backward compatible replacement for 2371627-2. DO NOT USE 2371627-2 to replace 2371627-3. 2371627-3 is EMC2 Compliant. ms13452 PART ADDED PER VCT RSPL AND IS EMC2 COMPLETE MS8336 12/08/2004
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required