Corduroy artifacts caused by CTL coil.
Symptom
Corduroy image artifacts only on HD CTL coil with overange errors in log. see long description See below for the errors. Mon May 18 14:10:08 2009 Error: 2270004 Host: signa2 Process: LxMxConsumer File: VrfTrace.cpp Line: 53 Server Name: ADS The VRF detected that MR Signal is too high (overrange) at channel: 0. The VRF is located at ICN1(ICN2), if ADS(ADS1) reports this error. Mon May 18 14:10:08 2009 Error: 2270058 Host: signa2 Process: LxMxConsumer File: VrfTrace.cpp Line: 89 Server Name: ADS The VRF detected that MR Signal is too high (overrange).
Error message
2248108 APS received data over range error from DRF, on chan= n. 2248108 TOiMR2 NSP APS:dataMoverTask 604 Exception Class: Idle Severity: Unknown Function: Acquisition/Pulse Generation: Acquisition PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 APS received data over range error from DRF, on chan = 2 Need to reduce receiver gain.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2377426-2 USAI 8CH CTL COIL
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.