Pular para o conteúdo principal
GE Healthcare · MRSigna224832115642221962822236252248355

Several images would not reconstruct Due to ICN.

Symptom

Some or All Images in the Last Scan Did Not Reconstruct Some or All Images in the Last Scan Did Not Reconstruct.

Error message

2248321 Communication Error. The callback for SCSM state 0x4 was not received in time. The DP server is not responding, the server might be hung. Please perform a TPS reset to recover.-> 2248321 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: VIBRANT Coil: 8Brst BL SMI Scan: 15642/6 Communication Error. The callback for SCSM state 0x4 was not received in time. The DP server is not responding, the server might be hung. Please perform a TPS reset to recover.->

2219628 The BAM Check Packet failed to reach the TPS Tue Jun 15 08:58:55 2004 Error: 2219628 Host: IRIS Process: scn File: SrxIPScreen.m Line: 1956 The BAM Check Packet failed to reach the TPS.

2223625 Send packet opcode:192 seqnum:2673 rev:2 type:1 length:36 req:SCAN_01_NOR resp:RECN_00_NOR status:0 failed. 2223625 Exception Class: Unknown Function: Software/Firmware: Host Communications Send packet opcode:192 seqnum:2673 rev:2 type:1 length:36 req:SCAN_01_NOR resp:RECN_00_NOR status:0 failed.

2248355 AGP to SCP State Error. SCT received a wavegen status 3 in a non-wavegen state 3 It may be necessary to perform a TPS reset. Error: 2248355 Host: hopgemr1a Process: NSP File: SCP:sct_mct

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.