Pular para o conteúdo principal
GE Healthcare · MRSigna22185372224667

Scan stopped due to an Infiniband backlog during data acquisition due to Image Compute Node.

Symptom

Scan stopped with Infiniband backlog during data acquisition errors Scan stopped with Infiniband backlog during data acquisition errors.

Error message

2218537 Scan stopped due to an Infiniband backlog during data acquisition 2218537 Mar 18 Fri 13:31:59 2218537 scn MsaScanMgr.m 3087 Scan stopped, Scan Stopped due to error :  Scan stopped due to an Infiniband backlog during data acquisition. Please try the scan again. Perform a TPS reset if the problem reoccurs..

2224667 Manual Prescan failed,Scan stopped due to an Infiniband backlog during data acquisition Thu Mar 17 19:06:32 2011 2224667 1 BDMR02 psc PscMainControl.cc 1392 ^MException Class: Unknown Function: Prescan: Host Manual Prescan Manual Prescan failed, Scan stopped due to an Infiniband backlog during data acquisition. Please try the scan again. Perform a TPS reset if the problem reoccurs.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5337894-3 Image Compute Node - HDxt 16CH GEN 3 5337894-3 Image Compute Node - HDxt 16CH GEN 3 Sun 4170
Area
MR