Pular para o conteúdo principal
GE Healthcare · MRSigna22538502541257

MRIQ Bright White Line in middle of Image and Gradient Hammer Test Failures Due to STIF Board.

Symptom

MRIQ Collapsed Image Bright White Line Down or Across Middle of The Image

Error message

2253850 Gradient Link Hammer test failed. The GP received an invalid X axis data value via the gradient data link. See long description for entire error. Sun Jun 19 00:22:58 2005 Error: 2253850 Host: lx-mr Process: diagCoord File: grad_hammer.c Line: 286 Diagnostic: Grad Hammer Iteration: 8 of infinite Action: 1 Task ID: 0x3d5c6c8 Time: 2541257 Object: SCP Test Task Gradient Link Hammer test failed. The GP received an invalid X axis data value via the gradient data link. Data Exp = 0007ffffH or 00080000H Data Rd = 00080010H Check the gradient data and clock fiber optic cable connections

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues