GE Healthcare · MRSigna2224727
Table Longitudinal Drive Problems Due to Gear Reducer 40 to 1 Gear Box.
Symptom
Intermittent Longitudinal Drive Problems Losing drive voltage intermittently at motor.
Error message
2224727 Unable to start prescan! Table not at scan plane Fri Nov 19 11:34:03 2004 Error: 2224727 Host: IRIS Process: psc File: PscMainControl.cc Line: 1296 Unable to start prescan! Table not at scan plane.
Applies to
- Models
- Signa
- Product
- 3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 46-282359P2 Gear 16 mm or 40-1 speed reducer Part of the longitudinal motor drive assembly in line with motor in rear pedestal. See long description 15 46–282359P2 2 SPEED REDUCE 1 40 TO 1 NON–MAGNETIC GEAR REDUCER Note: Part Description in FEMC is simply "GEAR 16 MM"
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.