GE Healthcare · CTLightSpeedVCT26013464512797026013464315000284000260100042
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Difficulty in positioning the cradle. Cradle will not drive normally and scan aborts with errors.
Error message
Error: 260134645 Cradle was stopped unexpectedly. Or cradle did not move for motion request. See long description Tue Nov 2 07:35:57 2010 260134645 tgp tgp_subsys gtcb_cmc_actions.cxx 678 Tue Nov 2 07:35:57 2010 Host : Table Sub-System Ermes # : 260134645 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_actions.cxx Line# : 678 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 . Cradle was stopped unexpectedly. Or cradle did not move for motion request. current position(um) : 127970 target position (um) : 0 target velocity (um/s): 5000 Error: 260134643 Cradle speed correction cannot cover such a large speed error See long description for full error Mon Mar 24 02:06:22 2008 Error: 260134643 Host: tgp Process: tgp subsys File: gtcb cmcmon actions const.cxx Line: 181 Mon Mar 24 02:06:22 2008 Host : Table Sub-System Ermes # : 260134643 Exception Class : Secondary Severity : Pri/Soft File : gtcb cmcmon actions const.cxx Line# : 181 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle speed correction cannot cover such a large speed error. target speed 0 um/s current speed error 15000 um/s corrected speed 284000 um/s
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required