GE Healthcare · CTBrightSpeedHiSpeed260118324150001799740
BrightSpeed Select, Delight Select or MERI Select
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118324 Unable to bring up rotor due to acceleration interval constraint. (This is a message, not an error - See Long Description) See the Long Description for full error code When you get this error look for previous errors in the log. This error is usually the effect of another error and not the root cause. (example: if a tube over temperature is the root cause, you will get this error if another scan is immediately attempted after the over temperature error. The cooling delay allows the rotor to cool between acceleration attempts, but is not the root cause of the problem) Wed Apr 20 14:45:38 2005 Error: 260118324 Host: orp Process: orp_subsys File: gen_exec.c Line: 3937 Wed Apr 20 14:45:37 2005 Host : On-Board Rotating Processor Ermes # : 260118324 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 3937 Function : XRAY GENERATION : Tube Rotor Control Scan Type : Helical Scan: 90/2/1 Unable to bring up rotor due to acceleration interval constraint. Remaining Time: 83 seconds RotState:0 Mode:0x3 RmExpD:15000 Accel:6960 (Rx:20 Last brake:6585 Decel:5950 EndExam:1799740 NoRxes:5950 Tmrs: Rotor Idle:0 Restore:0 Fil Idle:0 Restore:0)
Applies to
- Models
- BrightSpeed, HiSpeed
- Product
- BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5392132 Programmed Rotation Board - replaces 2409043, 2242440-11, 2214605-3 in Jedi Generator Auxiliary Module (This rotation brd. used on ALL HPower Systems and YMS Hispeed) See Long Description for Details. 5392132 Programmed Rotation Board - replaces 2409043, 2242440-11, 2214605-3 in Jedi Generator Auxiliary Module (This rotation brd. used on ALL HPower Systems and YMS Hispeed) Part #: 5392132 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FRU- Programmed Rotation V3 board FRU: Yes Status: Orderable Comments: ** FE Assist ** For P500D or Xray RAD, order part number 2409043. CT and Vascular continue to order 5392132. 212010438 3/1/11 17Aug2010 - Promoted for ECO 2104218 212006824 for 305024364. Part #: 2409043 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: FRU-PROGRAMMED ROTATION BOARD FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution