GE Healthcare · CTLightSpeed18316250029
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
183162 MA fault detected. Fault latch address see Long Description Sep 16 06:56:12 2009 Error: 183162 Host: OBC Process: Generator File: gen control.c 1.39 Line: 587 Function: XRAY GENERATION : Filament Selection Scan: 50029/9/15 Type: Axial Exception Level: Pri/Soft Time: 06:56:12:300 Log Series: 297 MA fault detected. Fault latch address = ffdf4d hex, Status = 0x87.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2217038 OBC backplane used on 2X/3X/4X systems See LD for details Part #: 2217038 Part Type: GP CONSUMABLES Part Weight: 3.19 LBR Repairable: No Returnable: No Hazardous Material: N Description: OBC BACKPLANE FRU: Yes Status: Orderable Parts Offered as: Current: 2217038 Harvested: Not Available Exchange: Not Available Comments: FLS - H2 GANTRY: COMMON TO MFG. MS26561 DSM 1/7/01
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.