LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
ECG Viewer is not receiving data from monitor when Gating BPM / ECG Trace button is clicked. Getting a pink pop-up box with ECG Viewer is not receiving data from monitor when they click on the Gating BPM / ECG Trace button with in the Fast Chest protocol. There are no errors posted in the messages box. . Getting a pink pop-up box with ECG Viewer is not receiving data from monitor when they click on the Gating BPM / ECG Trace button with in the Fast Chest protocol. There are no errors posted in the messages box. The information in the attention box is ECG Viewer is not receiving data from monitor. Viewer is closing. Please verify that the monitor is on and the ethernet cable is connected, then turn Gating OFF and ON to restart ECG Viewer
Error message
200610002 Communication error between EkgBroker and EkgViewer See Long Discription for Details Wed Dec 9 08:43:19 2009 Error: 200610002 Host: vct1 Process: EKGBroker File: EKGBroker.cxx Line: 292 EKGBroker could not communicate with EKGViewer. Message: Communication error between EkgBroker and EkgViewer. EkgViewer needs to be restarted. Error: 5
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2280889 Gantry Option Board GOB See Long Description for Details Part #: 2280889 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: GANTRY OPTION BOARD FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.