Symptom

"Scan hardware Stopped Scan" with beam tracking errors. "Scan hardware Stopped Scan" with beam tracking errors

Error message

Error: 260007456 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance. See LD for more details Error: 260007456 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance Mon Mar 20 16:36:22 2006 Error: 260007456 Host: DAS Process: DCB Firmware File: zaxis.c Line: 578 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50201/5/19 Type: Axial Exception Level: Pri/Soft Time: 16:36:22:380 Log Series: 83 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance. Tolerance : 200 umd Control Error : 319 umd Beam Side : A Side

Do Not Use - use on of the other two Error: 260101023 The (DCB or CCB) has notified the system of an error see long description for examples. Tue Dec 28 09:41:31 2004 Error: 260101023 Host: orp Process: orp_subsys File: rcib_erlog.c Line: 366 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260101023 Exception Class : Abort Severity : Primary File : rcib_erlog.c Line# : 366 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260007402 Notification level: 5 Sequence number: 1 The above err

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5140001-3 (replaces 5140001-2) Helios 3 collimator with notched filler bar and rating plate (used on 1.X-4.X) See Long Description for more info Part #: 5140001-3 Part Type: GP REPAIRABLES Part Weight: 115.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 3 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: ** FE Assist ** 1.x 2.x 3.x 4.x started with Ultras added minor quality improvements to the dash 2 and is backwards compatible (12/6/06 212001261) Part #: 5140001-2 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 2 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: This part is replacement for 2214768-2, please do not use stock of 2214768-2. ms13452 050516 Can use the 5140001-3 if no Stock of the -2 (9/26/05 212001261)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.