GE Healthcare · MROptima22246662269910227570831260660
Scanning intermittently interrupted with RF Hub errors due to the RF HUB CONTROL BD (RFCB).
Symptom
Scanning intermittently interrupted with RF Hub errors See Long Description Scanning intermittently interrupted, one or more RF Hub errors seen in error log:
Error message
2224666 Auto Prescan failed, RF Hub link is not ready 2224666 b3a psc ../PscMainControl.cc 1424 Exception Class: Unknown Function: Prescan: Host Autoprescan Auto Prescan failed, RF Hub link is not ready. 2269910 The CPLD watchdog timed out causing a reset. Board RF Hub Likely firmware error. 2269910 b3a NSP SCP:EclCpld 121 ^MException Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Fiesta Coil: Body 24 AA2 Scan: 328/28 The CPLD watchdog timed out causing a reset. Board: RF Hub Likely firmware error. 2275708 The RF Hub failed to recover CAN communication on wakeup. The RRX is reporting that RF Hub power is OK and clock is on. See Long Description Error: 2275708 Host: C-3TMRI2 Process: diagCoord File: RfHubCanHWO Line: 5792 Diagnostic: CFB Board Level Diagnostics Iteration: 1 of infinite Action: 1 Task ID: 0x2537bb0 Time: 31260660 Object: SCP Function: RF: RF Receive Communication Failure. The RF Hub failed to recover CAN communication on wakeup. The RRX is reporting that RF Hub power is OK and clock is on. Check RF Hub CAN cable connections.->
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Multicoil Bias Fault errors on multiple coils.
- Auto Prescan Failures and RF amp setup incomplete Errors Due to Defective RF amp.
- The system is unable to scan with any surface coil, get dynamic disable faults.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.