Pular para o conteúdo principal
GE Healthcare · MROptima22246662269910227570831260660

Scanning intermittently interrupted with RF Hub errors due to the RF HUB CONTROL BD (RFCB).

Symptom

Scanning intermittently interrupted with RF Hub errors See Long Description Scanning intermittently interrupted, one or more RF Hub errors seen in error log:

Error message

2224666 Auto Prescan failed, RF Hub link is not ready 2224666 b3a psc ../PscMainControl.cc 1424 Exception Class: Unknown Function: Prescan: Host Autoprescan Auto Prescan failed, RF Hub link is not ready.

2269910 The CPLD watchdog timed out causing a reset. Board RF Hub Likely firmware error. 2269910 b3a NSP SCP:EclCpld 121 ^MException Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Fiesta Coil: Body 24 AA2 Scan: 328/28 The CPLD watchdog timed out causing a reset. Board: RF Hub Likely firmware error.

2275708 The RF Hub failed to recover CAN communication on wakeup. The RRX is reporting that RF Hub power is OK and clock is on. See Long Description Error: 2275708 Host: C-3TMRI2 Process: diagCoord File: RfHubCanHWO Line: 5792 Diagnostic: CFB Board Level Diagnostics Iteration: 1 of infinite Action: 1 Task ID: 0x2537bb0 Time: 31260660 Object: SCP Function: RF: RF Receive Communication Failure. The RF Hub failed to recover CAN communication on wakeup. The RRX is reporting that RF Hub power is OK and clock is on. Check RF Hub CAN cable connections.->

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.