System won't scan with Loop Contactor error
Symptom
System won't scan with Loop Contactor error 183949 and 183921 Tech had completed the scout and had begun the exam when she started getting resumes.She ended the and restarted the exam but now could not scout. Error Log: Axial Error: 183949 File: ax error.c Method: No Method Line: 1106 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Scan: 4718/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 10:41:19:725 Log Series: 673 Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor. Axial Gantry Command/Status:0x9e08. Wed Apr 21 10:38:43 1999 Suite: STJ1 Host: STC Proc: Axial Error: 183921 File: ax error.c Method: No Method Line: 1106 Function: AXIAL CONTROL : Axial Loop Contactor interlocks Scan: 4717/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:38:39:730 Log Series: 659 Axial conditions at FAULT: velocity fb = -1 position fb = -55075 Formatted data: 8100 100 10 1 1 0 1 0 0 1 1 1064 346112 55075 18 -19 127 128 0x009e28
Error message
Error: 183949 Could NOT CLOSE Gantry (Axial) Loop Contactor See LD for full error CT01 Host: STC Proc: Axial Error: 183949 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor Another example: Tue May 24 08:41:21 2005 Suite: CT02 Host: STC Proc Axial Error: 183949 File: ax_error.c Method No Method 1114 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Scan: 65001/11/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:21:940 Log Series: 60 Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor. Axial Gantry Command/Status:0x9e28. EN 401 SR 442
Applies to
- Models
- HiSpeed, LightSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2114535 Right Side Gantry control panel - used on H1 gantry, HSA/CTi/QXi See LD for more info 2114535 Right Side Gantry control panel - used on H1 gantry, HSA/CTi/QXi 2114535-2 Left Side Gantry control panel - used on H1 gantry, HSA/CTi/QXi See LD for more info 2114535-2 Left Side Gantry control panel - used on H1 gantry, HSA/CTi/QXi Note: Rubber Overlay for the control panel is 2225299-2
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required