Pular para o conteúdo principal
GE Healthcare · MRSigna225230522523572252293

2252305 TR Driver Fault Detected Open Circuit Source of Fault is Body TR Due to Hybrid Splitter.

Symptom

Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.

Error message

2252305 TR Driver Fault Detected Open Circuit Source of Fault is Body TR. Thu Aug 5 11:00:38 2004 Error: 2252305 Host: fmri3te Process: NSP File: SCP:DCB Line: 1197 TR Driver Fault Detected - Open Circuit. Source of Fault is Body TR. Value: 16 mA Limit: 2400 mA->

2252357 Driver Module error occurred. 2252357 1 lx-mr NSP APS:aps_fastta 135 Driver Module error occurred.

2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2370221 3T DUAL BODY HYBRID aka 3T DUAL BODY T/R SWITCH ASM See Long Description In Renewal Parts, the part name is "3T Dual Body Hybrid". In FEMC the same part number has the part name "3T DUAL BODY T/R SWITCH ASM". Part # 2370221. This part is for the LCC300 short bore 3T both on G3 systems and Excite HD/HDx/HDxT. NOT compatible with the 3T94 long bore 3T.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues