Pular para o conteúdo principal
GE Healthcare · MRSigna224943111680822494321011620

Scan pauses intermittently with errors logged due to SCP3 board.

Symptom

Scan pauses intermittently with errors logged See Long Description Customer reports that sometimes scans pause. But when they press the scan button on the SCIM it continues the scan. Or sometimes it can not scan, but if they copy/paste the series and download it, the system can scan successfully. Errors are logged in the message log when this happens. See the cause concept and the associated Error concepts for the specific errors that are logged.

Error message

2249431 A PCI error was detected, but message queue send failed to the exception logging task see long description Error: 2249431 Host: GEMR_OC4 Process: diagCoord File: mgd820PciErr.c Line: 1378 Task ID: 0x3e2b6e0 Time: 116808 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A A PCI error was detected, but message queue send failed to the exception logging task; errno = 0x3D0002. Harrier Err Enable Reg: 0x00000000; Harrier Err Status Reg: 0x00000000. Harrier Err Address Reg: 0x00000000; Harrier Err Attribute Reg: 0x8000.

2249432 A PCI error occurred. The MCP820 received a Target Abort. See Long Description Error: 2249432 Host: EMR1 Process: diagCoord File: excLog.c Line: 1389 Task ID: 0x6eca5f0 Time: 1011620 Object: AGP Exception Class: Unknown Function: Software/Firmware/Config: N/A A PCI error occurred. The MCP820 received a Target Abort. Harrier Err Enable Reg: 0xFFF0E000; Harrier Err Status Reg: 0x40000000. Harrier Err Address Reg: 0x09016034; Harrier Err Attribute Reg: 0x0600. Additional information in /usr/g/service/log/vxPciLg-f6f77 and /usr/g/service/log/vxPciSum-f6fa4

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2294300-22 MGD SCP3 RoHS Compliant board 2294300-22 MGD SCP3 RoHS Compliant board
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues