Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2300090412000027182639020000274833928

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

SRU has gone down. SRU has gone down

Error message

Error: 230009041 The image did not complete reconstruction within the time allotted. See long description for complete error Tue Apr 15 19:45:47 2003 Error: 230009041 Host: smopct1 Process: queuemgr File: qmgr_queue_cntl.c Line: 2524 Function: Image Generation : SBC Processing Job Type: Retro Axial Scan: 414/4/1 Image: 414/4/5 The image did not complete reconstruction within the time allotted.

Error: 200002718 Image create Pixel data request failed See the long description for the full error code Thu Jun 30 23:03:01 2005 Error: 200002718 Host: ip01 Process: imagecreate File: ic imagecreate req.c Line: 247 Image : 26390/2/3 Recon Sequence Id : 15 Image create Pixel data request failed

Error: 200002748 Timeout waiting for pixel data response from icproxy See long description for a full sample error. Tue Aug 19 12:10:33 2008 Error: 200002748 Host: mrct2mid Process: imagecreate File: ic getpixeldata.c Line: 324 Image : 33928/2/83 Recon Sequence Id : 990 Timeout waiting for pixel data response from icproxy

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.