LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
ECG Viewer is not receiving data from monitor when Gating BPM / ECG Trace button is clicked. Getting a pink pop-up box with ECG Viewer is not receiving data from monitor when they click on the Gating BPM / ECG Trace button with in the Fast Chest protocol. There are no errors posted in the messages box. . Getting a pink pop-up box with ECG Viewer is not receiving data from monitor when they click on the Gating BPM / ECG Trace button with in the Fast Chest protocol. There are no errors posted in the messages box. The information in the attention box is ECG Viewer is not receiving data from monitor. Viewer is closing. Please verify that the monitor is on and the ethernet cable is connected, then turn Gating OFF and ON to restart ECG Viewer
Error message
Error: 260100083 Timeout Occurred Waiting for an External Gate Signal. Check Gating Hardware/Patient Connection See Long Description for complete error Tue Jan 22 13:34:24 2008 Error: 260100083 Host: tgp Process: tgp_subsys File: scExternalGate.c Line: 1030 Tue Jan 22 13:34:23 2008 Host : Table Gantry Processor Ermes # : 260100083 Exception Class : Abort Severity : Primary File : scExternalGate.c Line# : 1030 Function : No System Function Reported Scan Type : Axial Scan: 1318/2/1 Scan Aborted. A Timeout Occurred Waiting for an External Gate Signal. Check Gating Hardware/Patient Connection.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5198566 cable-gantry - EKG PRO See Long Discription for Details Part #: 5198566 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: cable-gantry - exg pro FRU: Yes Status: Orderable 5112819-2 ECG, 3-LEAD, LOW-NOISE PATIENT CABLE, 3m LONG see LD 5112819-2 ECG, 3-LEAD, LOW-NOISE PATIENT CABLE , 3m LONG Part #: 5112819-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ECG, 3-LEAD, LOW-NOISE, PATIENT CABLE, 3m LONG FRU: Yes Status: Orderable
- Area
- CT