Symptom

Scanner failed to initiate scan with warning the subsystem firmware had been shutdown Scanner failed to initiate scan with a red warning that the subsystem firmware had been shutdown.

Error message

Error: 244572 UnExpected Event received sender:orp_subsys event:51004 time:0 See long description for full sample error message. Mon Aug 11 08:36:32 2008 Error: 244572 Host: h0150ct1 Process: fwmgr File: fwmgr_events.c Line: 9449 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 UnExpected Event received sender:orp_subsys event:51004 time:0

Error: 245226 Scanmgr requested hardware reset. Scanning hardware might be down, or the communication with the firmware might have failed. See LD for complete error Sun Sep 10 05:29:21 2006 Error: 245226 Host: cpro Process: scanmgr File: scanmgr_lib.cxx Line: 1450 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 1084000 Scanmgr requested hardware reset. Scanning hardware might be down, or the communication with the firmware might have failed.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues