GE Healthcare · CTLightSpeedBrightSpeedHiSpeed23001781200050026011821410132
LightSpeed 5.X (80/100kW)
Symptom
Scanner Hardware Fault with high voltage spits. Scanner Hardware Fault with high voltage spits
Error message
Error: 230017812 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500 See Long Desciption for complete error Mon May 30 05:41:49 2005 Error: 230017812 Host: cxi1 Process: accs File: collSweepProcessing.c Line: 491 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500 Error: 260118214 (60-0317) Spit Ratio error See Long Description for complete error. Fri Dec 16 23:26:55 2005 Error: 260118214 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Fri Dec 16 23:26:54 2005 Host : Jedi Ermes # : 260118214 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Axial Scan: 10132/2/1 (60-0317)Spit Ratio error. PPC board has detected the rate of tube spits during exposure has reached the limit (see theory of operation, software section).^C Z6,Z7= 4h,0h
Applies to
- Models
- LightSpeed, BrightSpeed, HiSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2266521-5 HV Tank (replaces 2266521-4 and 2266521-3 - backward compatable) used on LS 7.X, LS 5.X and LightSpeed RT. See Long Description for details 2266521-5 HV Tank (replaces 2266521-4 and 2266521-3 - backward compatable) used on LS 7.X, LS 5.X and LightSpeed RT. Dash Tanks includes the EMI Shielding needed for G-Das Part #: 2266521-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Part #: 2266521-4 Part Type: GP CONSUMABLES Part Weight: -5499945.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Part #: 2266521-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 200.84 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Comments: added per ecr 2005924, as replacement for 2266521 and 266521-2. ms13452 040923 **If there is no Local Stock for the -3 part, you may order the -4 version which is backward compatible** ms0262 060815 *** 2ND OPINION NOTES *** Run the No-load KV test to verify HV tank is the error.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required