Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26010721850053260112009260100095

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

No AC power to the gantry. Gantry display is blank

Error message

Error: 260107218 Gantry Drive Switch found DISABLED See LD for complete error Wed Jun 11 08:09:02 2008 Error: 260107218 Host: tgp Process: tgp_subsys File: amc_mon.c Line: 221 Wed Jun 11 08:09:02 2008 Host : Table Gantry Processor Ermes # : 260107218 Exception Class : Secondary Severity : Pri/Soft File : amc_mon.c Line# : 221 Function : AXIAL CONTROL : Axial Control Scan Type : Axial Scan: 50053/1/2 Gantry Drive Switch found DISABLED.

Error: 260112009 Emergency Off button has been pressed See LD for full error Fri May 20 15:06:56 2005 Error: 260112009 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr.c Line: 640 Fri May 20 15:06:56 2005 Host : Table Gantry Processor Ermes # : 260112009 Exception Class : Operator Abort Severity : Secondary File : uif_table_tilt_mgr.c Line# : 640 Function : SYSTEM POWER CONTROL : Emergency Stop Scan Type : None/Unknown Scan: 0/0/0 Emergency Off button has been pressed

Error: 260100095 HVDC Switch is Disabled See Long Discriptions for Detail Sat Apr 2 10:10:08 2011 Error: 260100095 Host: tgp Process: tgp_subsys File: scGantryInMail.c Line: 1067 Sat Apr 2 10:10:07 2011 Host : Table Gantry Processor Ermes # : 260100095 Exception Class : Secondary

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues