The Home Flag on the Longitudinal Drive Belt was not passing through the Home Flag Detect Optical Coupler.

Symptom

No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.

Error message

2224842 Table motion was stopped because it moved differently than expected. See Long Description 2224842 CNSJOMR2 NSP SCP:SRI 100 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Fiesta Coil: HD BodyUpper Scan: 769/9 Table motion was stopped because it moved differently than expected. This can occur if the table was moved by hand. Last Velocity Commanded: 100 mm/sec Measured Cradle Velocity: -0 mm/sec-

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
46-282436P1 CARRIAGE TRAVEL SENSOR FLAG If the screws are broke or missing you will need two of them to hold the flag onto block. 46–265486P8 SCREW If the screws are broke or missing you will need two of them to hold the flag onto block. 46–265486P8 SCREW.