Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000683110821

LightSpeed 4.X - Octane

Symptom

Scanner Hardware stopped scan message with GANTRY CONTACT switch FAULT during Rotation of the Gantry. CALL Service - Emergency Stop of the Axial Drive. See long description for details Scanner Hardware stopped scan message with GANTRY CONTACT switch FAULT during Rotation of the Gantry. CALL Service - Emergency Stop of the Axial Drive. This indicates that something may be loose inside gantry and hit one of the rotate fault switches.

Error message

Error: 260006831 Axial Motor Control (AMC) logic detected GANTRY CONTACT switch FAULT during Rotation of the Gantry. See LD for complete error Sat Jan 1 16:17:39 2005 260006831 2 STC Axial amc isr.c 789 Function: AXIAL CONTROL : axial control Scan: 10821/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 16:17:39:370 Log Series: 849 Axial Motor Control (AMC) logic detected GANTRY CONTACT switch FAULT during Rotation of the Gantry. CALL Service - Emergency Stop of the Axial Drive. (AMC State:8 Cmd:0x15b30008 CurVel:811 (.001 sec/rev)).

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-195120G21 HV CABLES 3 PIN 7 FT See LD for details Part #: 46-195120G21 Part Type: GP CONSUMABLES Part Weight: 3.19 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLES 3 PIN 7 FT FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.