GE Healthcare · CTLightSpeedVCT26010102350444260007402260007453260001512260007353
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
"Scanner Hardware Stopped Scan" with CAM or Filter errors
Error message
Do Not Use - use on of the other two Error: 260101023 The (DCB or CCB) has notified the system of an error see long description for examples. Tue Dec 28 09:41:31 2004 Error: 260101023 Host: orp Process: orp_subsys File: rcib_erlog.c Line: 366 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260101023 Exception Class : Abort Severity : Primary File : rcib_erlog.c Line# : 366 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260007402 Notification level: 5 Sequence number: 1 The above error is a catch all for many different notifications and sequences. Other examples: Host : Das Ermes # : 260101023 Exception Class : No System Function Reported Scan Type : Helical Scan: 8564/2/1 The DCB has notified the system of an error condition which prevents scanning. See log for further information: Error message code: 260007453 Notification level: 5 Sequence number: 17 Wed Sep 22 11:21:43 2004 Error: 260101023 orp orp_subsys rcib_erlog.c 363 Wed Sep 22 11:21:41 2004 Host : Das Ermes # : 260101023 Exception Class : Abort Severity : Primary Fil
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5357424 (includes 5130001 Collimator and shipping box) 32 slice VCT Failsafe Collimator FRU only. Not for 64 slice - see SN11091210a See long description for details. There are two part numbers for ordering the VCT Collimator. The old one is the 5130001. This part number is still available for ordering as of Oct 2009. The other part number, 5357424, is a FRU assembly that includes the 5130001 and a new shipping container. Part #: 5357424 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT Collimator FRU FRU: Yes Status: Orderable Comments: **FE ASSIST** This collimator is only to be ordered for PRO32 systems.(212024490 Nov 16,2011) Otherwise:CT Collimator for Gantry 5126093-x. If Your Gantry is 5124069-x or 5232083-x must order Collimator 5410597. Refer to SN11091210a (212001261 Sept 22 2011) Part #: 5130001 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VCT Failsafe Collimator Asm FRU: Yes Status: Orderable Parts Offered as: Current: 5130001 Harvested: Not Available Exchange: Not Available Comments: Added per ecr: 2014028, this is Failsafe collimator and is replacement for: 2383390. Do NOT use 2383390 as replacement for 5130001 as it is not forward compatible, ms13452 050419
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required