GE Healthcare · CTLightSpeed26010011050070
"Scan hardware Stopped Scan" with PDU errors
Symptom
"Scan hardware Stopped Scan" with PDU errors "Scan hardware Stopped Scan" with PDU errorss
Error message
Error: 260100110 HVDC KSS Contactor is NOT Closing based on the Relay Status. Timemout:2000 msec Status:0. -OR YOU MAY SEE- The PDU Kss Contactor has NOT closed in the alloted time. Timeout:300 msec Timer:0 msec Contactor Status: Open Error: 260100110 HVDC KSS Contactor is NOT Closing based on the Relay Status. Timemout:2000 msec Status:0.Tue Jul 27 18:25:12 2004 Error: 260100110 [insite@hvllct01 ~]$ /usr/g/insite/bin/viewlog Host: tgp Process: tgp subsys File: scHvdcMgr.c Line: 307 Tue Jul 27 18:25:13 2004 Host : Table Gantry Processor Ermes # : 260100110 Exception Class : Abort Severity : Primary File : scHvdcMgr.c Line# : 307 Function : SYSTEM MONITORING : HVDC Monitor Scan Type : Scout Scan: 4949/1/1 HVDC KSS Contactor is NOT Closing based on the Relay Status. Timemout:2000 msec Status:0. This error is flagged by Table Gantry Processor on the SYSTEM MONITORING : HVDC Monitor. ---------Similar error below but for Axial Scan------------ Tue Feb 7 07:50:11 2006 Error: 260100110 Host: tgp Process: tgp_subsys File: scHvdcMgr.c Line: 328 Tue Feb 7 07:50:11 2006 Host : Table Gantry Processor Ermes # : 260100110 Exception Class : Abort Severity : Primary File : scHvdcMgr.c Line# : 328
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT4 WB O G4 G HPT WB O M4 G HPT LightSpeed RT4 with GRE and MDAS4 or GDAS4 gantry At GE Medical Systems, we’re pioneering exciting CT technology for Radiation Therapy planning. Introducing the newest addition to our LightSpeed® series – LightSpeed RT, the first multi-slice, wide-bore CT scanner with the largest field of view in the industry. With a generous 80 cm gantry opening, the LightSpeed RT bore is 10 cm larger than conventional scanners. This provides complete flexibility for patient positioning freedom – even for the most demanding oncology needs. And the 65 cm variable field of view delivers 15 additional centimeters of visual coverage, allowing clinicians to see a patient’s entire anatomy for precise radiation therapy planning. Position and precision Thin, multi-slice imaging is the extra bonus – enabling respiratory gating, providing precise target delineation and delivering higher resolution digitally reconstructed radiographs (DRRs) than ever before. All adding up to the ability for you to deliver what really matters most – better cancer care for your patients. Plus, the LightSpeed RT scanner is available with the Exact™ couch insert, enabling you to consistently replicate patient immobilization for treatment. With LightSpeed RT, the future of your Radiation Oncology department is wide open. Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2334820-3 replaces 2334820-2 PDU control board Used by BSD,HD,VCT, RT,LS,CJ and PET(BSD) with NGPDU or NGPDU-4 See LD for details 2334820-3 replaces 2334820-2 PDU control board Used by BSD,HD,VCT, RT,LS,CJ and PET(BSD) with NGPDU or NGPDU-4 Part #: 2334820-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: NGPDU control board FRU: Yes Status: Orderable Comments: 30 Aug 2011 promoted for ECR2122276.305024364
- Area
- CT
Full cause & step-by-step resolution