LightSpeed 5.X (80/100kW)
Symptom
When trying to move cradle in or out of the gantry, it moves about ½ an inch and then stops. See long description for details. The table will move up and down. When trying to move cradle it or out of the gantry, the cradle moves about ½ an inch and then stops.
Error message
Error: 260125003 Table Communication has been established. Fri May 20 15:14:49 2005 Error: 260125003 Host: tgp Process: tgp_subsys File: ta_func.c Line: 246 Fri May 20 15:14:34 2005 Host : Table Gantry Processor Ermes # : 260125003 Exception Class : Soft Severity : Pri/Soft File : ta_func.c Line# : 246 Function : SYSTEM COMMUNICATIONS : Table Communications Scan Type : None/Unknown Scan: 0/0/0 Table Communication has been established. Error: 260115190 Cradle encoder fault See the long description for the full error Wed May 25 17:35:00 2005 Error: 260115190 Host: tgp Process: tgp_subsys File: mtcb_cmc_isrfunc.c Line: 163 Wed May 25 17:35:00 2005 Host : Table Sub-System Ermes # : 260115190 Exception Class : Soft Severity : Pri/Soft File : mtcb_cmc_isrfunc.c Line# : 163 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle encoder fault . Error: 244633 Table reports that Table Communication becomes down. Wed May 25 17:35:00 2005 Error: 244633 Host: ct01 Process: fwmgr File: fwmgr_events.c Line: 7301 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Table reports that Table Communication becomes down. Error: 2601250
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2342727-4 : 5.X Cradle encoder. This part replaces the 2342727-3 and 2342727-2 This is an absolute position encoder with no pot. See long description for details Displaying Change-Up Data: Part "2342727-2" and "2342727-3" changed to 2342727-4 Part #: 2342727-4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FRABA OCD H-Power String Encoder FRU: Yes Status: Orderable Comments: Added by ecr: 2020264. Please continue to use stock of 2342727-3, until depleted. -3 and -4 are mfg alts of one another. ms13452 Revised per to Asset confirmation Comments: Added per ECR2011883 to replaced the 2342727-2 The 5.x table design incorporates Absolute encoders used for cradle and elevation, instead of incremental encoders. The absolute encoder has the capability to generate a unique binary 'word' for each encoder shaft position, allowing the encoder to provide positional information instantly upon power-up, unlike incremental encoders. Therefore, potentiometers are NOT needed for cradle or elevation encoder circuits.
- Area
- CT