GE Healthcare · MRSigna226117622611802228179

Short Circuit in 16 Channel Switch caused failure in SRI.CoilID Not Working for Any Coil CoilID Light Never Illuminates.…

Short Circuit in 16 Channel Switch caused failure in SRI.CoilID Not Working for Any Coil CoilID Light Never Illuminates.Replace 16 Channel Switch and SRI AFTER checking this Long Description wqr.22611…

Symptom

CoilID Not Working for Any Coil CoilID Light Never Illuminates See Long Description. System is being upgraded from 3T94 to 3T HDx. Coil ID isn't working for any coil and there are no error messages when plugging in a coil. The CoilID lamp or light never comes on. We expect to see either green or red (in case of failure) but it never lights up at all. However we see the "coil presence" LED illuminate at the 16 channel switch when plugging in a coil. We do see the error message shown below when booting up the system and when running a [TPS Reset]. Tue Feb 5 18:56:45 2008 2261176 2 rmno NSP SCP:SriAutoConfig 831 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power. Error: ECHO->

Error message

2261176 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. 2261176 tempmr NSP SCP:SriAutoConfig 796 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power. 2261176 tempmr NSP SCP:SriAutoConfig 796 SRI to 16 Channel Switch Communication Error:An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power.

2261180 SRI to IRFM1 Communication Error: An error occurred reading the serial ID on the IRFM1 Error: 2261180 Host: MRI01 Process: NSP File: SCP:SriAutoConfig Line: 864 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 SRI to IRFM1 Communication Error: An error occurred reading the serial ID on the IRFM1. Please check the connection between the SRI and IRFM1

2228179 TPS detected NSP (host) timeout Tue Sep 20 14:17:14 2005 Error: 2228179 Host: lx-mr Process: NSP File: APS:hs watchdog check Line: 10 TPS detected NSP

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
SecondOpinion 1.5T Excite HD 16 Channel Switch 5109645 See comments in resolution concept for second opinion requirements before ordering this part 5105270 Scan Room Interface 3 Assembly - (SRI 3)
Area
MR