LightSpeed 4.X - Octane

Symptom

Scanner Hardware Stopped Scan HEMRC Errors See Long Description for more information. Scanner Hardware Stopped Scan message appears during Helical Scanning with the following Errors: 185553 (The HEMRC drive has detected a Line Loss - Alarm status: 20H. This occurs when the input power to the drive falls below 85% of the nominal Bus voltage) Possible causes: X-Ray tube stator , HEM-IT, HEMRC power supply 183318 (The HEMRC output current is out of 45.0 percent WARNING tolerance), Monitored state = 2 (2=Run state) Expected current = 2.748 A Actual current = 4.097 A. 183319 (The HEMRC output current is out of 55.0 percent MAXIMUM tolerance), Monitored state = 2 (2=Run state) Expected current = 2.748 A Actual current = 4.653 A. 183316 (The HEMRC failed to reach the Run speed in the allotted time), and 185548 (The HEMRC operating frequency is below the minimum value) Actual frequency: 106.973 Minimum frequency limit: 130.000 At Frequency signal: Inactive Expected: Inactive Rotor State = Transition

Error message

Error: 185557- hemrc serial link is down. Hemrc_+12V failure detected. Possible causes: HEMRC fuse (F5), Interconnects, HEMRC AC Drive, HEMRC Control Board See Long Description for complete error Mon Jul 24 08:41:32 2006 Error: 185557 Host: OBC Process: Generator File: scanbusDriver.c 1.5 Line: 243 Function: XRAY GENERATION : Tube Rotor Control Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:28:915 Log Series: 31 Could not send HEMRC message in the allotted time. HEMRC serial link is down. HEMRC_ISO_+12V Failure detected. Possible causes: HEMRC fuse (F5), Interconnects, HEMRC AC Drive,

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2145832 HEMRC I/F Ctrl Board - This board is used in many different places in CTi (with Performix Tube) through 4.X LightSpeed See Long Description for the many places this board is used. 2145832 HEMRC I/F Ctrl Board - This board is used in many different places in CTi (with Performix Tube) through 4.X LightSpeed CTi (with Performix Tube Only), and QXi use this board in the Tube Rotor Controller (HEMRC). NOT used in AXIAL control. 2.X, 3.X, and 4.X use this board in the Tube Rotor Controller (HEMRC) and also for axial control (also HEMRC) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.