Auto prescan fails with signal too small on head coil due to Poor Seating of Quad Head Coil Cables and/or Quick Disconne…
Auto prescan fails with signal too small on head coil due to Poor Seating of Quad Head Coil Cables and/or Quick Disconnect Box.
Symptom
auto prescan fails with signal too small on head coil auto prescan fails with signal too small on head coil.
Error message
2224666 Auto Prescan failed, signal too small. Please try Manual Prescan. Fri Jan 23 14:39:55 2004 2224666 lx-mr psc PscMainControl.cc 1271 Auto Prescan failed, signal too small. Please try Manual Prescan.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2148308-2 HEAD COIL QUICK DISCONNECT (NORMAL POLARITY) 1.5T
- Area
- MR
Full cause & step-by-step resolution