GE Healthcare · MRSigna2224666221853722259352225695
The operator's knee hit the scim cable connection beneath the console tabletop.
Symptom
Scan Aborts with MDS Link Error
Error message
2224666 Auto Prescan failed, SPI-MDS link is not working. 2218537 Scan stopped, the Emergency Stop was pressed or PDU 24V power supply has failed. 2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. 2225935 The PAC does not respond. Check PAC power and fiber optic connections, run PAC diags. This error condition adversely affects gated scans only. 2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. This error condition adversely affects gated scans only. 2225695 MDS link failure. An MDS peripheral indicated an error but did not send an error code. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. See Long Description Timer: 50 MDS peripheral address: 0x00
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.