Pular para o conteúdo principal
GE Healthcare · MRDiscoverySigna2224580

Center Frequency was off.Auto prescan Fails with Signal Too Small message on all coils.Correct Center Frequency wqr.Disc…

Center Frequency was off.Auto prescan Fails with Signal Too Small message on all coils.Correct Center Frequency wqr.Discovery MR750.2224580 Auto Prescan failed.

Symptom

Auto prescan Fails with Signal Too Small message on all coils Auto prescan fails with Signal Too Small message on all coils

Error message

2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan

Applies to

Models
Discovery, Signa
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues