Fiber Optic Link Errors Due to IRF I/O Board.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

TPS is not responding caused by RRF-DIF Fiber Optic Link Error

Error message

2254012 RRF-DIF Fiber Optic Link Error on the Master RRF-DIF board. Tx Ready for Data signal is in an Invalid State.

2254012 RRF-DIF Fiber Optic Link Error on the Master RRF-DIF board. Rx Signal Detect signal is in an Invalid State. Tue May 14 19:20:37 2002 2254012 1 bay12b_oc0 NSP SCP:RRF-DIF 630 RRF-DIF Fiber Optic Link Error on the Master RRF-DIF board. Rx Signal Detect signal is in an Invalid State.

2247160 RRF Fiber Cable Link Not Ready See Long Description Fri Jan 25 09:23:14 2008 Error: 2247160 Host: FHMR01 Process: diagCoord File: Irf.cpp Line: 272 Diagnostic: MUX BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 51176 Object: APS Test Task RRF FIBER CABLE LINK NOT READY The fiber link between the RRF and MGD chassis is not in a ready state after hardware initialization. This will prevent the MGD from controlling the RRF chassis. Try TPS_reset. Run RRF diagnostics.-> Extended Error: The MGD does not like the signal coming over the fiber cable connected from the IRF_IO board in the back of the MGD chassis to the DIF board in slot 10 in the front of the RRF chassis. Possible cause(s): RRF-DIF Board, GigaLink Fiber Cable, IRF I/O Board, IRF Board

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2274196 IRF I/O Board See Long Description. Per FEMC part information, Description of part number 2274196 is an INTERFACE and REMOTE RF TRANSITION BOARD. Part Comments as of 2/8/2008 are as follows; FLS - INTERFACE & REMOTE RF TRANSITION ADDED PER RSPL. USED ON PHOENIX MGD. MSA8429 05/14/02 ROUTE TO MR FOR CONSUMABLE COLLECTION PER B. HAHLEN, 7-11-02, A. MASCH THIS PART WAS ADDED AS PART OFTHE EXCITE NPI FOR MR. 9/01/02 MS24445 BH FLS - FE ASSIST 08OCT2002 (MS2045) ADDED TO MR DISTRIBUTION LIST FOR PARTS TRACKING.
Area
MR

Full cause & step-by-step resolution