Pular para o conteúdo principal
GE Healthcare · CTLightSpeed245206

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Unable to scan with a message on screen that says: Collimator/DAS reset failed.

Error message

Error: 245206 Software has detected an invalid or mistached DAS type. This will prevent t he system from scanning. Sun Sep 5 08:11:02 2004 Error: 245206 Host: ct01 Process: scanmgr File: validate_request.cxx Line: 341 Software has detected an invalid or mistached DAS type. This will prevent t he system from scanning.

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P75 15A 250 V DAS power fuse. See Long description for details P/N 46-170021P75 15A 250 V DAS power fuse Part #: 46-170021P75 Part Type: GP CONSUMABLES Part Weight: 0.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: 015.000A 0250V 3AB SLOW UL FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues