GE Healthcare · MRSigna2254702
Buttons on Magnet Enclosure Control Panels do not work due to swapped cables
Symptom
Buttons on Magnet Enclosure Control Panels do not work No table functions, no bore light, no alignment light, no fan control. Estop works normally.
Error message
2254702 PAC DIAGNOSTIC TIMEOUT. The PAC did not return a result to the SPU for the last PAC diagnostic test in the allotted 8 second time. Try running PAC datapath tests or other PAC diagnostics. Extended Error: Verify that the PAC module has power. Run SPU-->STIF and SPU-->PAC loopback tests to see if the MGD chassis can talk to the PAC module. Check fiber optic connections between the STIF and the PAC. Possible cause(s): PAC
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.