Buttons on Magnet Enclosure Control Panels do not work due to swapped cables

Symptom

Buttons on Magnet Enclosure Control Panels do not work No table functions, no bore light, no alignment light, no fan control. Estop works normally.

Error message

2254702 PAC DIAGNOSTIC TIMEOUT. The PAC did not return a result to the SPU for the last PAC diagnostic test in the allotted 8 second time. Try running PAC datapath tests or other PAC diagnostics. Extended Error: Verify that the PAC module has power. Run SPU-->STIF and SPU-->PAC loopback tests to see if the MGD chassis can talk to the PAC module. Check fiber optic connections between the STIF and the PAC. Possible cause(s): PAC

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR