Will Not Scan With RF Errors.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2225710 RF Interface (RFI) Mismatch Fault. The Dump Resistor in the RFI has overheated. see long Error: 2225710 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3064 RF Interface (RFI) Mismatch Fault. The Dump Resistor in the RFI has overheated. Possible causes: improper calibration, missing or improper cables between RFI and amplifier(s), defective RFI or defective TELI RF Amplifier(s).

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2194013 1.0T RF INTERFACE SCALEABLE RF (RFI module) 1.0T RFI module 2194013 RF INTERFACE SCALEABLE RF.
Area
MR

Full cause & step-by-step resolution