GE Healthcare · MRSigna2248653
Intermittent Warnings of high Patient Comfort Temperature Readings due to Gradient Chiller.
Symptom
Intermittent Warnings of high Patient Comfort Temperature Readings Intermittent Warnings of high Patient Comfort Temperature Readings.
Error message
E02 Error on AirSys Chiller Display. The AirSys Medicool chiller control display reads Error E02 as the error code for Low pressure (digital) or high water level. Active if low pressure (suction pressure) is below 0.6Bar or when the water level in the tank is below the high level. (Likely a coolant leak) See Long Description. The AirSys Medicool chiller control display reads Error E02 as the error code for Low pressure (digital) or high water level. Active if low pressure (suction pressure) is below 0.6Bar or when the water level in the tank is below the high level. (Likely a coolant leak) The Air Sys Medicool 10.0 P6 R407C chiller User Manual is here . (see page 35) 2248653 New scans can be initiated, but the patient comfort level is still warmer than normal. 2248653 1 b4a NSP SCP:BoreTempSriHWO 1009 New scans can be initiated, but the patient comfort level is still warmer than normal.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.