GE Healthcare · MRSigna22554852223880
Applications boot fails due to corrupt software.
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
2255485 Local Lookup of NSP failed. Packet not sent. rpc status is -1 See long description 2255485 LxMxConsumer cmTransmit.cpp 428 Server Name: DP. Error: Local Lookup of NSP failed. Packet not sent. rpc status is -1 2223880 Reset TPS was not successful. Board Failure Status APS Fail, SCP Fail, AGP Fail, VRE Pass. ErrorNo 9 See Long Description Error: 2223880 Host: OUMCGEMR1ET Process: NSP File: TardisReset.cc Line: 491 Exception Class: Unknown Function: Host: NSP Reset TPS was not successful. Board Failure Status: APS: Fail, SCP: Fail, AGP: Fail, VRE: Pass. ErrorNo: 9.-> Extended Error: The MGD failed to start a TPS Reset. Either the serial connection to the MGD STIF Board is not working or the Ethernet connection from the Host to the MGD Hub is not working. Run the APS, AGP, and SCP Host Data path ping tests to check the Ethernet connections. Run the APS, AGP, and SCP Board Level Diagnostics to check the SCSI Serial Expansion. The STIF Board serial connection is not tested. If both sets of diagnostics pass but yet the TPS Reset still fails then please closely inspect the serial connection between the SCSI Serial Expansion and the MGD STIF Board. APS, AGP, and SCP Board Level Dia
Applies to
- Models
- Signa
- Product
- 3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.