Known Software issue with 15M4B coding error.

Symptom

When scanning DWEPI a pop up message indicates /usr/g/caldir/delay.espb.xyz could not be found... using defaults. Continue? see long description Customer advisory pop up message occurs when performing a DWI Sequence in the 8ch Head coil. Message pop-up: /usr/g/caldir/delay.espb.xyz could not be found... using defaults. Continue? the customer can accept and continue but defaults will be used. This is an indication that the group delay calibrations were not completed properly.

Error message

pop-up advisory /usr/g/caldir/delay.espb.xyz could not be found... using defaults. Continue? Message pop-up: /usr/g/caldir/delay.espb.xyz could not be found... using defaults. Continue?

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR