Short Circuit in 16 Channel Switch caused failure in SRI.
Symptom
CoilID Not Working for Any Coil CoilID Light Never Illuminates See Long Description. System is being upgraded from 3T94 to 3T HDx. Coil ID isn't working for any coil and there are no error messages when plugging in a coil. The CoilID lamp or light never comes on. We expect to see either green or red (in case of failure) but it never lights up at all. However we see the "coil presence" LED illuminate at the 16 channel switch when plugging in a coil. We do see the error message shown below when booting up the system and when running a [TPS Reset]. Tue Feb 5 18:56:45 2008 2261176 2 rmno NSP SCP:SriAutoConfig 831 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power. Error: ECHO->
Error message
2261176 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. 2261176 tempmr NSP SCP:SriAutoConfig 796 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power. 2261176 tempmr NSP SCP:SriAutoConfig 796 SRI to 16 Channel Switch Communication Error:An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion 1.5T Excite HD 16 Channel Switch 5109645 See comments in resolution concept for second opinion requirements before ordering this part 5105270 Scan Room Interface 3 Assembly - (SRI 3)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.