GE Healthcare · MRSigna2254253

2254253 3T RF Amplifier PA Filament under Voltage (fault 38) due to blown fuse in MKS 3.0T RF Amplifier.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2254253 3T RF Amplifier PA Filament under Voltage (fault 38). The PA Filament supply output was below specification(15V - 5%). 2254253 NSP SCP:../rf_nb_contro 1460 memp 8HRBRAIN protocol:DQA_8ch_Brain/1 3T RF Amplifier PA Filament under Voltage (fault 38). The PA Filament supply output was below specification(15V - 5%). Possible Causes: Defective supply or short on supply output. If the fault persists, replace the RF amplifier

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2336064-8 3T CORONA MKS RF AMPLIFIER FUSE KIT 3T CORONA RF AMPLIFIER FUSE KIT 2336064-8 3T CORONA RF AMPLIFIER FUSE KIT 2336064-8