Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18240718241039139

LightSpeed 4.X - GRE

Symptom

Cradle Drive Motion was prohibited by hardware Interference switches

Error message

Error: 182407 Velocity error detected when trying to find the Index Pulse Sun Aug 29 13:25:59 2004 Error: 182407 Host: ETC Process: Elevation File: CmdMotion.c 1.35 12/23/93 Line: 174 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 13:25:59:690 Log Series: 34 HW/SW error: Velocity error detected when trying to find the Index Pulse. Hardware Inhibit Switches: 0xc0d8 Elev Direction: 0 (0 = Up 1 = Down). INFO: Probable Problem could be 180V Power is LOW.

Error: 182410 HW/SW error: Motion was prohibited by hardware Limit switches Thu Aug 2 21:40:36 2007 Error: 182410 Host: ETC Process: Elevation File: emcControl.c 1.23 11/13/92 Line: 168 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 39139/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:40:37:175 Log Series: 764 HW/SW error: Motion was prohibited by hardware Limit switches. Hardware Inhibit Switches 0xc2d8 Current Position 4943 (.1 mm) Dir: 1 (0 = Up 1 = Down). Characterization Limits 5005 (.1 mm) 250 (.1 mm)

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.