GE Healthcare · MRSigna2225713
Unable to scan with Power Monitor Faults due to HN Bulkhead Connector or Body Heliax Feed Through Connector.
Symptom
Unable to scan with Power Monitor Faults
Error message
2225713 Power Monitor 1 Fault. RF Transmit Power has Reached Safety Limit Allowed for Scan Parameters and Patient Weight. See long description Thu Apr 15 11:50:18 2004 Error: 2225713 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_pm_log_error Line: 100 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 255 pulse width: 07 duty cycle: 02 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan. This could also occur when power cycling SSM.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.