Pular para o conteúdo principal
GE Healthcare · MR2225551

Multi Coil TR Driver Fault.

Symptom

Unable to Scan with Multi Coil TR Driver Faults

Error message

2225551 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 1. See Long Description Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225551 File: PMH.m 1.29 Method: spi_tr_mon Line: 6230 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 1. Check cable to multi-coil board or check for an open pin diode.

Applies to

Product
HF OpenSpeed HFO1 HFO2 and HFO3 Only.
Software
HFO3
Component
2264743-9 0.7T SHOULDER ARRAY COIL
Area
MR