GE Healthcare · MRSigna22247092225551
Multicoil open circuit fault due to damaged CTL cable.
Symptom
Unable to Scan with Multi Coil TR Driver Faults
Error message
2224709 Low signal detected. Please check the signal level for receiver x or receiver x+1 while using Manual Prescan. Error: 2224709 Host: NMI_Langhorn Process: NSP File: TPS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver 1 (or receiver 2 while using Manual Prescan). Note that for 3 coil scans, one coil will cause a low signal detected message to occur. Error: 2224709 Host: NMI_Langhorn Process: NSP File: TPS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver (X + 1) while using Manual Prescan). Note that for 3 coil scans, one coil will cause a low signal detected message to occur. 2225551 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 1. See Long Description Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225551 File: PMH.m 1.29 Method: spi_tr_mon Line: 6230 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 1. Check cable to multi-coil board or check for an open pin diode.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2133085 MULTICOIL CABLE, 1.5T FOR TELEDYNE CTL ARRAY 2133085 MULTICOIL CABLE, 1.5T FOR TELEDYNE CTL ARRAY
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.