Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260116088

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Can't perform Scout Scan. Can't perform Scout Scan Please go to the error concept and check all that apply The customer may see Scanner Hardware Stopped Scan message in the operator console area 1 Use the following script and search the message log at time of Scout failure to get the correct errors. [insite@erct ~]$ /usr/g/insite/bin/viewlog 2 Type those errors in the psdb search box separate by spaces 3 Go to the error concept and check the appropriate errors

Error message

Error code:260116088 the hardware watchdog was not running See long description for details Error code:260116088 the hardware watchdog was not running Wed Dec 1 11:21:32 2010 260116088 2 tgp tgp_subsys sys_watchdog_task.c 361 Wed Dec 1 11:21:31 2010 Host : Table Gantry Processor Ermes # : 260116088 Exception Class : Safe State Severity : Secondary File : sys_watchdog_task.c Line# : 361 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The hardware watchdog was not running when it should be Scanning Hardware Reset is needed to recover

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.