Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260006500129605127751201

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

Gantry will not rotate or scan, alignment lights will not turn ON

Error message

Error: 260006500 The axial AC Drive reported fault code: F5 - Overvoltage Fault. Please see long description The first error is Error 260006500 The axial AC Drive reported fault code: F5 - Overvoltage Fault. 1296051277 0 1 Wed Jan 26 14:14:37 2011 260006500 2 STC Axial AX2fault.c 196 Function: AXIAL CONTROL : axial control Scan: 51201/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 14:14:34:305 Log Series: 440 The axial AC Drive reported fault code: F5 - Overvoltage Fault The Drive internal DC bus voltage has exceeded 810V. Possible failure of axial I/F Board Chopper Control or excessive motor regeneration from AXMotor during braking.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2145832 HEMRC I/F Ctrl Board - This board is used in many different places in CTi (with Performix Tube) through 4.X LightSpeed See Long Description for the many places this board is used. 2145832 HEMRC I/F Ctrl Board - This board is used in many different places in CTi (with Performix Tube) through 4.X LightSpeed CTi (with Performix Tube Only), and QXi use this board in the Tube Rotor Controller (HEMRC). NOT used in AXIAL control. 2.X, 3.X, and 4.X use this board in the Tube Rotor Controller (HEMRC) and also for axial control (also HEMRC) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.