GE Healthcare · MRSigna2254352225434722612302254342
RF Amp Not Ready with Error Codes 2254347, 2254342 or 2254352 due to SRFD2 RF Amplifier.Will Not Scan With RF Errors.Sec…
RF Amp Not Ready with Error Codes 2254347, 2254342 or 2254352 due to SRFD2 RF Amplifier.Will Not Scan With RF Errors.SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 UPDATED Dec 2 2010 wqr.Si…
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2254352 RF Amplifier Interface Serial Link Failure Sun Nov 5 12:07:40 2006 Error: 2254352 Host: HORIMR01 Process: NSP File: SCP:../rf_nb_serial Line: 1530 RF Amplifier Interface <-> RF Amplifier serial link failure. The serial link appears to be broken. Check the RF Amp power and the serial link. 2254347 RF Amplifier Interface RF Amp serial link failure. RF AIF failed to receive a packet transmission echo on 2nd try. If link is broke, a separate error message will be logged. If not, a spurious error occurred. Timer: 5080 Expired time: 40 Tx by 2254347 PAC3T NSP SCP:../rf_nb_serial 1947 RF Amplifier Interface RF Amp serial link failure. RF AIF failed to receive a packet transmission echo on 2nd try. If link is broke, a separate error message will be logged. If not, a spurious error occurred. Timer: 5080 Expired time: 40 Tx byte: M 2254347 PAC3T NSP SCP:../rf_nb_serial 1947 RF Amplifier Interface <-> RF Amp serial link failure. RF AIF failed to receive a packet transmission echo on 2nd try. If link is broke, a separate error message will be logged. If not, a spurious error occurred. Timer: 5080 Expired time: 40 Tx byte: M 2261230 The SCP received an unknown frequency identifier fro
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.