Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

While Booting system will stay in "waitting for application start up" interface see long description System will stay in "waitting for application start up" interface

Error message

/var/log/messages startMon: VERB startup failed see long description insite@HENDCTDISCOVERY1 log$ cat messages.1 | grep VERB Aug 17 06:05:11 HENDCTDISCOVERY1 logger: verbpower: Powering up VERB Aug 17 06:17:06 HENDCTDISCOVERY1 startup: startMon: VERB startup failed Aug 17 13:07:45 HENDCTDISCOVERY1 logger: verbpower: Powering up VERB Aug 17 13:19:23 HENDCTDISCOVERY1 startup: startMon: VERB startup failed

Error: 0 VERB startup failed! See Long Description Sun Oct 3 05:29:07 2010 Error: 0 Host: ct04 Process: Svc_Notepad File: Notepad.c Line: 44 This message was added by the OPERATOR to report on a problem: VERB startup failed!

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5313765 Cable, Ethernet CAT5e, 1.5 Meter See long description Part #: 5313765 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Cable, Ethernet CAT5e, 1.5 Meter FRU: Yes Status: Orderable Comments: . . Per HDVCT RSPL, 212008898, 7/15/08.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues